Animas 2020 Guía de usuario Pagina 11

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 116
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 10
Page 11
1
Medical Procedures
Procedure Patient is Pumper Healthcare Professional/
Technician is Pumper
Pacemaker / AICD
Disconnect pump and leave outside room during
insertion of device and reprogramming. Infusion
set can remain in place.
Follow normal safety practices and
proceed to the designated safe
area while each X-ray is taken and
during reprogramming.
EKG
No need to disconnect. No need to disconnect.
Cardiac Catheter
Disconnect pump and leave outside room during
procedure.
Follow normal safety practices and
proceed to the designated safe area
while each X-ray is taken.
Nuclear Stress test
Disconnect pump and leave outside room during
scan. Can remain connected during injection of
radioisotope.
Follow normal safety practices and
proceed to the designated safe area
while each X-ray is taken.
Colonoscopy
No need to disconnect. No need to disconnect.
Laser Surgery
The pump and infusion set can be worn;
however some lasers can create interference and
cause the pump to alarm.
The pump and infusion set can
be worn; however some lasers can
create interference and cause the
pump to alarm.
General anesthesia
Determination based on what medical
equipment is being used in the procedure.
Determination based on what
medical equipment is being used in
the procedure.
For procedures not included above or on the chart on the following page, please call
(in advance of your procedure) 1-877-YES-PUMP.
Vista de pagina 10
1 2 ... 6 7 8 9 10 11 12 13 14 15 16 ... 115 116

Comentarios a estos manuales

Sin comentarios