Changing the Battery
Each time you change the battery
A full rewind and prime sequence is required.• See this section, Priming the Pump and
Infusion Set
The Insulin on Board calculation starts over at zero•
The Combo bolus returns to the factory set default duration and split•
Review your basal program settings•
Low Battery Warning means battery life will end in minimum 1 hour.
1. Use a coin to unscrew the battery cap with a counter-clockwise motion.
2. Check your battery cap for damage such as cracks or missing threads, and be sure the
colored (yellow) o-ring fits securely and is not torn or damaged.
3. Check the vent hole on the top of the battery cap to be sure it is clear of debris. This
vent maintains pressurization while preventing water from entering the compartment.
The battery cap should be replaced every six months or sooner if the o-ring or cap is
damaged or the vent is clogged. See Chapter 13, Care and Maintenance.
4. Insert the Energizer® Lithium L91 AA (1.5V) battery into the battery compartment with
the positive (+) end going in first.
5. Replace the cap by turning clockwise until you cannot see the o-ring. Then slowly
tighten the cap until flush.
6. Each time you change the battery, your Animas® 2020 will run a series of self-tests which
will last a few seconds. The Animas logo will be displayed briefly. The pump will give a beep
to alert you to verify (or change) the time/date, language and battery type.
Yellow o-ring
Battery Cap Vent Hole
Energizer
®
Lithium L91 AA battery