7
6. The Home screen is displayed and shows your Temp Basal is active, the percentage
change, the duration and how much time is left. When the duration of time is complete,
your pump will automatically resume the active basal program.
Note: If you turned on the Temp Basal sound in Setup, the pump will beep once every
30 minutes to remind you of Temp Basal status.
Canceling a Temporary Basal Program
1. From Basal Menu, select Temp Basal. Press OK.
2. Details of the current active Temp Basal program will be displayed. Scroll up to Cancel
and press OK.
Your previously active basal program will be activated and the Home screen will be displayed
to show the current rate per hour of the active basal program.
Note: If you Suspend your pump while a Temp Basal program is active, the Temp Basal program
will be canceled and an Alert
screen will notify you that the Temp Basal program has been
canceled. This Alert is displayed once and gives an audible tone once (if you turned on Alert
sounds in Setup). Temp Basal is also canceled when you change the battery and/or prime.