4
Contrast Button
Pressing this button adjusts the contrast of your display. There are three contrast levels:
Dim, Default and Bright. To preserve battery life, your pump display will Auto-dim when
a button is not pressed for half the time your display time-out is set. While in Auto-dim
mode, you can restore to the default contrast level you have set by pressing the Contrast
button on top of the pump. Pressing a function button while in Auto-dim mode will restore
the default contrast level as well as perform the function of the key. To adjust contrast
during a Call Service alarm, you must use the contrast button. See Chapter 11, Advanced
Features, Setup Advanced 4 – Contrast Adjustment and Display Timeout Setting.
Note: When viewing the pump display in bright sunlight, it is recommended you shade
the screen or move to a shady area for best visibility.
Audio Bolus/ezBolus™
This button allows you to program a bolus without looking at the pump, by using audible
tones to confirm programming and delivery*. If you choose not to activate the Audio Bolus
feature, this button provides a shortcut to the Normal Bolus screen.
*When you first use the Audio Bolus feature, you should always look at the screen to confirm
correct programming until you are comfortable with using audio feedback to program
a bolus. See Chapter 11, Advanced Features, Setup Advanced 1 – Audio Bolus Feature.
Battery Cap/Vent
This cap unscrews easily with a coin to replace and secure your battery. There is a yellow
o-ring around the cap, which prevents water from entering. The battery cap also is equipped
with a built-in vent to allow air to enter the pump to maintain pressurization but prevent
water from entering. Replace the cap by turning clockwise until you cannot see the o-ring.
Then slowly tighten the cap until flush. See Chapter 5, Changing the Battery.
Primary Vent
This vent is part of the redundant vent safety
system, which allows air inside the pump to
maintain equalized pressure but prevents water
from getting inside.
Under no circumstances should you introduce
any kind of sharp object into the vent openings
to clean them. Doing so will compromise the
pump’s waterproof capabilities. If at any time
you suspect the vent openings are clogged,
replace the battery cap or call Animas for questions about the Primary Vent.
Battery Cap/Vent
Primary Vent
Cartridge Cap