Animas 2020 Guía de usuario Pagina 57

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 116
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 56
Page 57
11
Setup Advanced 4 – Language Setup, Display Timeout, Contrast and Battery Type
This screen allows you to:
Select a different language•
Set the length of time your display stays on before timing out to save battery life•
15, 30, 45 or 60 seconds
Select a contrast setting•
Select Lithium (recommended) or Alkaline battery type•
1. Scroll to the desired field.
2. Press OK to change to flashing cursor for Edit mode.
3. Use the Up/Down arrows to change to desired setting.
4. Press OK when setting is made.
To move to the next Setup Advanced screen, scroll to the bottom of the screen and highlight
Next. Press OK.
To return to the Home screen, scroll to Home and press OK.
Contrast Button
Pressing the button on the top of your pump adjusts the contrast. There are three contrast levels:
Dim, Default and Bright. To preserve battery life, your pump will Auto-dim when a button is not
pressed for half the time your display time out is set. While in Auto-dim mode, you can restore
to the default contrast level you have set by pressing the Contrast button
. Pressing a function
button while in Auto-dim mode will restore the default contrast level as well as perform the
function of the key.
If in Call Service Alarm mode, you must use the Contrast button to
restore the default contrast level.
To return contrast setting to original factory default, press the Up Arrow and Contrast button at
the same time. When the wordContrast is displayed on the screen, press any button to return
to the default contrast setting.
Note: When viewing the pump display in bright sunlight, it is recommended you shade
the screen or move to a shady area for best visibility.
Setup Advanced 5 – Auto-OFF Feature
This screen allows you to set the pump to automatically suspend basal delivery and sound
an alarm if no buttons are pressed in a user-selected number of hours. This feature can be
used as a safeguard in case the user is unconscious.
1. Scroll to the desired field.
2. Press OK to change to flashing cursor for Edit mode.
3. Use the Up/Down arrows to change to desired setting.
4. Press OK when setting is made.
To move to the next Setup Advanced screen, scroll to the bottom of
the screen and highlight Next. Press OK.
To return to the Home screen, scroll to Home and press OK.