Alert: Low BG
Cause BG entry below 3.9 mmol/L.
Effect Requires user confirmation to continue.
Message Displayed until confirmed.
Action Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: High BG
Cause BG entry above 13.9 mmol/L.
Effect Requires user confirmation to continue.
Message Displayed until confirmed.
Action Press OK to confirm.
Beeps/Vib
User selected, one time and each time manually
awakened until confirmed. No progression.
Alert: Clear Program Basal Segments
Cause Clear command selected from Basal Options screen.
Effect Requires user confirmation to continue.
Message
Displayed until confirmed or until pump goes
to sleep.
Action Select Clear Program or Basal Options.
Beeps/Vib User selected, one time.
Alert: Basal Program Display Change
Cause
Changing display of basals from 4 to 1 but program 1
is not currently active.
Effect Requires user confirmation to continue.
Message Displayed until confirmed or until pump goes to sleep.
Action Press OK to confirm.
Beeps/Vib User selected, one time.
Alerts, Warnings and Alarms